Chapter 8: Troubleshooting
11 - Printer Timeout
Message: “Printer has timed out. Check that the printer is on and selected.”
Explanation: The printer did not respond and is not ready to print. Confirm that the printer’s selection light is
green (you may have to press the
SEL
or
Online
button on the printer).
15 - ID Board Warning
Message: “A problem has been detected with the ID Board. Please contact GE.”
Explanation: A communication error with the ID board has occurred. Turn the Analyzer off and on and restart
analysis. If the warning persists, contact Technical Support.
16 - Flow Started
Message: “Sample stream flow restored.”
Explanation: The flow of sample water to the Analyzer has been restored after a system-reported interruption.
This message automatically clears the related 63 - Flow Stop error from the screen, and replaces the E (Error)
icon at the top of the screen with a W (Warning) icon.
18 - UV Lamp Warning
Message: “The UV lamp needs to be replaced.”
Explanation: Based on the history of usage, the Analyzer estimates that the UV lamp has expired. Order a UV
lamp now. Continuing to use the expired UV lamp can result in incomplete oxidation and poor measurements.
21 - Tubing Warning
Message: “The tubing needs to be replaced.”
Explanation: Based on the history of usage, the Analyzer estimates that the sample pump tubing has expired.
Order new pump tubing from GE Analytical Instruments now.
22 - Resin Bed Warning
Message: “The resin bed needs to be replaced.”
Explanation: Based on the history of usage, the Analyzer estimates that the resin bed has expired. Order a new
resin bed from GE Analytical Instruments now.
23 - Autozero Warning
Message: “TOC Autozero cancelled -- sample too variable. Autozero will be retried at next scheduled time.”
Explanation: The Analyzer was unable to perform the TOC Autozero due to fluctuations in the IC or TC levels of
the sample stream. The Analyzer waits until the next scheduled time and attempts to perform the Autozero
again. If this warning persists, there may be an issue with the water system.
25 - History Data Warning
Message: “A problem has been corrected in the history data. Contact GE if the warning persists.”
Explanation: The Analyzer has detected a problem reading or writing to the data history memory; as much of the
data history as possible is reconstructed. If the problem continues, contact GE Analytical Instruments technical
support.
26 - System Restart Warning
Message: “The system has experienced an error and has automatically restarted. Contact GE if the warning
persists.”
Explanation: The Analyzer has restarted for a reason other than a power failure. If the problem continues, there
may be a problem with the Analyzer and you should contact GE Analytical Instruments technical support.
Summary of Contents for Sievers 500 RL
Page 8: ...GE Analytical Instruments 2010 8 of 226 DLM 74001 04 Rev A ...
Page 10: ...GE Analytical Instruments 2010 10 of 226 DLM 74001 04 Rev A ...
Page 36: ...GE Analytical Instruments 2010 36 of 220 DLM 74001 04 Rev A ...
Page 66: ...GE Analytical Instruments 2010 66 of 226 DLM 74001 04 Rev A Chapter 3 Installation ...
Page 152: ...GE Analytical Instruments 2010 152 of 226 DLM 74001 04 Rev A Chapter 7 Maintenance ...
Page 170: ...GE Analytical Instruments 2010 170 of 226 DLM 74001 04 Rev A Chapter 8 Troubleshooting ...
Page 178: ...Appendix A GE Analytical Instruments 2010 178 of 186 DLM 74001 04 Rev A ...
Page 185: ...Notes GE Analytical Instruments 2010 185 of 186 DLM 74001 04 Rev A 186 ...
Page 186: ...Notes GE Analytical Instruments 2010 186 of 186 DLM 74001 04 Rev A 186 ...