
GE Analytical Instruments ©2010
126 of 226
DLM 74001-04 Rev. A
Chapter 6: Calibration and System Protocols
11. Press the
Exit
.
12. If you have a Super
iOS
system, and selected the
After
or the
Before and After
option, the Analyzer will
prompt you to remove the standards, and to insert the rinse cartridge or vials into the Super
iOS
system
to continue the Rinse activity. Take out the rinse cartridge or vials when completed.
13. If you have a standard
iOS
System, remove the standards and slide the
iOS
door closed.
Figure 29: The Accuracy, Precision, and Verification Summary Screen, Part 1
Figure 30: The Accuracy, Precision, and Verification Summary Screen, Part 2
Relative Standard Deviation (RSD) =
Standard Deviation
Measured TOC Concentration
--------------------------------------------------------------------------
100
Summary of Contents for Sievers 500 RL
Page 8: ...GE Analytical Instruments 2010 8 of 226 DLM 74001 04 Rev A ...
Page 10: ...GE Analytical Instruments 2010 10 of 226 DLM 74001 04 Rev A ...
Page 36: ...GE Analytical Instruments 2010 36 of 220 DLM 74001 04 Rev A ...
Page 66: ...GE Analytical Instruments 2010 66 of 226 DLM 74001 04 Rev A Chapter 3 Installation ...
Page 152: ...GE Analytical Instruments 2010 152 of 226 DLM 74001 04 Rev A Chapter 7 Maintenance ...
Page 170: ...GE Analytical Instruments 2010 170 of 226 DLM 74001 04 Rev A Chapter 8 Troubleshooting ...
Page 178: ...Appendix A GE Analytical Instruments 2010 178 of 186 DLM 74001 04 Rev A ...
Page 185: ...Notes GE Analytical Instruments 2010 185 of 186 DLM 74001 04 Rev A 186 ...
Page 186: ...Notes GE Analytical Instruments 2010 186 of 186 DLM 74001 04 Rev A 186 ...