Chapter 4: Basic Analyzer Operation
Limits. Limits will trigger if any of the TOC or conductivity tests fail. TOC Limit only if any of the TOC tests fail, and
Cond Limit only if any of the conductivity tests fail.
Selecting the Program Language
If you have loaded a second program language into the Analyzer, you can switch between the two languages by
following these steps:
1. Select the
Maintenance
tab.
2. Press the
Advanced
button.
3. Press the
Advanced Setup
button.
4. Press the
Select Language
button. Two buttons will display; the choices depend on which languages
you have loaded into the Analyzer. Select the desired language to continue.
Note that if only one program language is loaded, the
Select Language
button does not display.
Configuring the Flow Sensor
In configurations with an
iOS
or Super
iOS
, the Analyzer automatically detects the presence of the sample
stream when it is working in On-Line mode. When the water flow is interrupted, the Analyzer issues an error and
stops analysis. To change the
Flow Sensor
setting, follow these steps:
1. Select the
Maintenance
tab.
2. Press the
Advanced
button.
3. Press the
Advanced Setup
button.
4. Press the
Flow Sensor
button.
5. By default, the Flow Sensor is set to
On
. To change the Flow Sensor Status, press the
Flow Sensor
button and then select
On
or
Off
.
Press the
Flow Restart
button and set it to
On
or
Off
. Setting the
Flow Restart
to
On
means the Analyzer
will automatically begin analysis again after a temporary stoppage of sample flow, if it was taking
Note
If you change the program language by mistake and need help navigating the menus to
switch back to your native language, see “Problems Changing the Program Language” on
page 165.
Note:
Every time the Analyzer is stopped and re-started in On-Line mode, the Flow Sensor
automatically resets to On, to protect the Analyzer.
Summary of Contents for Sievers 500 RL
Page 8: ...GE Analytical Instruments 2010 8 of 226 DLM 74001 04 Rev A ...
Page 10: ...GE Analytical Instruments 2010 10 of 226 DLM 74001 04 Rev A ...
Page 36: ...GE Analytical Instruments 2010 36 of 220 DLM 74001 04 Rev A ...
Page 66: ...GE Analytical Instruments 2010 66 of 226 DLM 74001 04 Rev A Chapter 3 Installation ...
Page 152: ...GE Analytical Instruments 2010 152 of 226 DLM 74001 04 Rev A Chapter 7 Maintenance ...
Page 170: ...GE Analytical Instruments 2010 170 of 226 DLM 74001 04 Rev A Chapter 8 Troubleshooting ...
Page 178: ...Appendix A GE Analytical Instruments 2010 178 of 186 DLM 74001 04 Rev A ...
Page 185: ...Notes GE Analytical Instruments 2010 185 of 186 DLM 74001 04 Rev A 186 ...
Page 186: ...Notes GE Analytical Instruments 2010 186 of 186 DLM 74001 04 Rev A 186 ...