Chapter 8: Troubleshooting
4. Press the
button to print the list.
You also can print or export a full list of warnings and errors that includes the time each event occurred, even
after you have removed items from the Warnings/Errors list. To do so, follow these steps:
1. Select the
Maintenance
tab.
2. Press the
Advanced
button.
3. Press the
Warn/Error Out
button.
4. Press either the
Print History
or
Export History
button.
If you select
Export History
, you will be prompted to select the port for export. Press either
Serial
or
USB
to export the list of warnings and errors.
A complete list of warnings and errors is shown below.
Warnings and Error Messages
5 - UV Lamp Warning
Message: “The estimated lamp life is less than 15 days. Order a new UV lamp.”
Explanation: Based on the history of usage, the Analyzer estimates that only 10% of the UV lamp life remains.
Order a new UV lamp from GE Analytical Instruments now to avoid down time when the lamp expires.
6 - Tubing Warning
Message: “The estimated tubing life is less than 15 days. Order new tubing.”
Explanation: Based on the history of usage, the Analyzer estimates that only 10% of the sample pump tubing life
remains. Order new pump tubing from GE Analytical Instruments now to avoid down time when the tubing
expires.
7 - Resin Bed Warning
Message: “The resin bed life is less than 15 days. Order a new resin bed.”
Explanation: Based on the history of usage, the Analyzer estimates that only 10% of the resin bed life remains.
Order a new resin bed from GE Analytical Instruments now to avoid down time when the resin bed expires.
8 - User List Warning
Message: “The user list is nearly full. Archive the user list before adding more users.”
Explanation: DataGuard allows a maximum of 100 users. This warning displays when 90 user accounts have
been created.
9 - History Data Warning
Message: “The history data is nearly full. Archive the history data before taking more measurements.”
Explanation: The data history can accommodate approximately 33,000 entries (about 90 days of usage). This
warning displays when the data history is 90% full (when approximately 31,941 entries have been saved).
10 - Audit Trail Warning
Message: “The audit trail is nearly full. Archive the audit trail before taking more measurements.”
Explanation: The audit trail can accomodate approximately 1,200 entries. This warning displays when the audit
trail is 90% full (when approximately 1,080 entries have been saved).
Summary of Contents for Sievers 500 RL
Page 8: ...GE Analytical Instruments 2010 8 of 226 DLM 74001 04 Rev A ...
Page 10: ...GE Analytical Instruments 2010 10 of 226 DLM 74001 04 Rev A ...
Page 36: ...GE Analytical Instruments 2010 36 of 220 DLM 74001 04 Rev A ...
Page 66: ...GE Analytical Instruments 2010 66 of 226 DLM 74001 04 Rev A Chapter 3 Installation ...
Page 152: ...GE Analytical Instruments 2010 152 of 226 DLM 74001 04 Rev A Chapter 7 Maintenance ...
Page 170: ...GE Analytical Instruments 2010 170 of 226 DLM 74001 04 Rev A Chapter 8 Troubleshooting ...
Page 178: ...Appendix A GE Analytical Instruments 2010 178 of 186 DLM 74001 04 Rev A ...
Page 185: ...Notes GE Analytical Instruments 2010 185 of 186 DLM 74001 04 Rev A 186 ...
Page 186: ...Notes GE Analytical Instruments 2010 186 of 186 DLM 74001 04 Rev A 186 ...