Chapter 4: Basic Analyzer Operation
1. Select the
Setup
tab (see Figure 11).
2. If you are running the Analyzer with a continuous water source, select an On-Line mode:
• Select
On-Line
to measure TOC continuously and return measurement results every 6 minutes.
• Select
On-Line Timed
to display a single TOC measurement at specific times of the day. Press the
Configure
button and specify the hour and minute for the first measurement; then, specify the time
interval for successive measurements (
30
Minutes
,
1 Hour
,
4 Hour
,
8 Hour
,
24 Hour
). At each
specified interval, the Analyzer displays data for the measurement that occurred at that time.
• Select
On-Line Averaged
to return the averaged value of the last 10 TOC measurements over a
specific time interval. Press the
Configure
button and specify the hour and minute for the first
measurement, then specify the time interval. At each specified interval, the Analyzer displays data
reflecting the average TOC value since the last specified interval. Also press the
Threshold
button to
specify a TOC level above which the Analyzer will switch to
On-Line
mode for the remainder of the
interval. When the threshold is exceeded, the Analyzer gives the average value for the interval thus
far and then returns measurements every 6 minutes.
3. If you are sampling from a vial in the Analyzer’s
iOS
or Super
iOS
System, press the
Grab
button. Follow
the instructions under “Grab Mode Measurement Settings” on page 70 for configuring Grab mode. Grab
mode is not available for configurations with the Sample Inlet Block.
4. Press the
Menu
button to exit and save your settings.
Figure 11: The Setup Tab
Grab Mode Measurement Settings
When selecting Grab mode (available on configurations with an
iOS
or Super
iOS
), you need to specify the
number of replicate measurements and rejects as follows:
1. On the
Setup
tab, press the
Configure Mode
button.
Summary of Contents for Sievers 500 RL
Page 8: ...GE Analytical Instruments 2010 8 of 226 DLM 74001 04 Rev A ...
Page 10: ...GE Analytical Instruments 2010 10 of 226 DLM 74001 04 Rev A ...
Page 36: ...GE Analytical Instruments 2010 36 of 220 DLM 74001 04 Rev A ...
Page 66: ...GE Analytical Instruments 2010 66 of 226 DLM 74001 04 Rev A Chapter 3 Installation ...
Page 152: ...GE Analytical Instruments 2010 152 of 226 DLM 74001 04 Rev A Chapter 7 Maintenance ...
Page 170: ...GE Analytical Instruments 2010 170 of 226 DLM 74001 04 Rev A Chapter 8 Troubleshooting ...
Page 178: ...Appendix A GE Analytical Instruments 2010 178 of 186 DLM 74001 04 Rev A ...
Page 185: ...Notes GE Analytical Instruments 2010 185 of 186 DLM 74001 04 Rev A 186 ...
Page 186: ...Notes GE Analytical Instruments 2010 186 of 186 DLM 74001 04 Rev A 186 ...