
GE Analytical Instruments ©2010
69 of 226
DLM 74001-04 Rev. A
Chapter 4: Basic Analyzer Operation
3. The Status area displays information about the current operation mode, the Analyzer’s current status
(Initializing, Idle, Analyzing, etc.), a countdown timer showing the time remaining in the current analysis,
and the repetition and rejection count, if applicable.
Taking TOC Measurements
After you have set the analysis mode and have configured the mode settings (as described in “Setting the
Analyzer Mode” on page 69) you are ready to begin taking TOC measurements. Simply press the
Start Analysis
button to begin.
The Menu Screen
The
Menu
screen provides you with access to all Analyzer settings and configuration options. Note that if
Password protection or DataGuard is enabled, some options may be available only to certain users. See
Figure 22 on page 102 for a menu map that indicates the required User Level to access various settings.
The
Menu
screen contains the five main configuration categories; press a tab to display the available options:
•
Setup
— Set the Analyzer mode and change parameters such as the number of repetitions and the
number of rejects.
•
Data
— Display and export historical data stored in RAM.
•
I/O
— Configure the input and output of data from the Analyzer.
•
Maintenance
— Display information about the Analyzer’s consumables levels, system information, and
error messages. Also perform calibration and system suitability operations.
•
DataGuard
— Configure the access settings and user passwords for your Analyzer.
From any tab under the
Menu
screen, you can press the
Start Analysis
button to initiate TOC measurements.
Setup
Options on the
Setup
tab let you specify preferences and basic operation settings for the Analyzer configuration.
If the sample conditions for your application remain relatively constant, you should seldom need to configure
the settings here.
Setting the Analyzer Mode
The Sievers 500 RL On-Line TOC Analyzer can operate in up to four different modes, depending on your
configuration. To set the Analyzer mode, follow these steps:
Summary of Contents for Sievers 500 RL
Page 8: ...GE Analytical Instruments 2010 8 of 226 DLM 74001 04 Rev A ...
Page 10: ...GE Analytical Instruments 2010 10 of 226 DLM 74001 04 Rev A ...
Page 36: ...GE Analytical Instruments 2010 36 of 220 DLM 74001 04 Rev A ...
Page 66: ...GE Analytical Instruments 2010 66 of 226 DLM 74001 04 Rev A Chapter 3 Installation ...
Page 152: ...GE Analytical Instruments 2010 152 of 226 DLM 74001 04 Rev A Chapter 7 Maintenance ...
Page 170: ...GE Analytical Instruments 2010 170 of 226 DLM 74001 04 Rev A Chapter 8 Troubleshooting ...
Page 178: ...Appendix A GE Analytical Instruments 2010 178 of 186 DLM 74001 04 Rev A ...
Page 185: ...Notes GE Analytical Instruments 2010 185 of 186 DLM 74001 04 Rev A 186 ...
Page 186: ...Notes GE Analytical Instruments 2010 186 of 186 DLM 74001 04 Rev A 186 ...