GE Analytical Instruments ©2010
80 of 226
DLM 74001-04 Rev. A
Chapter 4: Basic Analyzer Operation
For example, a successful measurement would output as follows:
Mode values (Field 1) can be On-Line, Timed (On-Line Timed), Averaged (On-Line Averaged), or Grab. For
information on Analyzer modes, see “Setting the Analyzer Mode” on page 69. In a warning or error condition,
some values may not be output.
After field 9, the Analyzer outputs a carriage return, followed by a line feed.
If configured to a 400 ES format, refer to the
“Series 400ES, 400DI, 2244OL TOC Analyzer Operations and
Maintenance Manual.
Issuing Serial Commands to the Analyzer
You can use a PC running a communications program, such as HyperTerminal, to issue commands to the
Analyzer via a serial (RS-232) connection or Ethernet (port 23). For information on establishing a connection,
consult “Using HyperTerminal” on page 179 in the Appendix. The Analyzer accepts the following serial
commands:
RUN
Start analysis
STP
Stop analysis
When the Analyzer receives one of these commands successfully, it returns
ACK
to the communications
program on the PC. If the command fails or if a bad command is sent, the Analyzer returns
NAK
.
Configuring Printer Settings
You can change the way the Analyzer formats output sent to the printer.
1. Select the
I/O
tab.
2. Press the
Printer
button.
3. Press the
Printer
button and choose
Epson
,
Seiko
, or
Citizen
, depending on the type of printer you
have. If no printer is attached to the Analyzer, choose
No Printer
. If you want to try using a printer made
by a different manufacturer, select
Citizen
, as it offers the most generic compatibility.
4. Set
Header Frequency
to
First Page
or
All Pages
to specify when a header will print. The header
includes basic information about the Analyzer, including the firmware version, Analyzer serial number,
and the current date.
Summary of Contents for Sievers 500 RL
Page 8: ...GE Analytical Instruments 2010 8 of 226 DLM 74001 04 Rev A ...
Page 10: ...GE Analytical Instruments 2010 10 of 226 DLM 74001 04 Rev A ...
Page 36: ...GE Analytical Instruments 2010 36 of 220 DLM 74001 04 Rev A ...
Page 66: ...GE Analytical Instruments 2010 66 of 226 DLM 74001 04 Rev A Chapter 3 Installation ...
Page 152: ...GE Analytical Instruments 2010 152 of 226 DLM 74001 04 Rev A Chapter 7 Maintenance ...
Page 170: ...GE Analytical Instruments 2010 170 of 226 DLM 74001 04 Rev A Chapter 8 Troubleshooting ...
Page 178: ...Appendix A GE Analytical Instruments 2010 178 of 186 DLM 74001 04 Rev A ...
Page 185: ...Notes GE Analytical Instruments 2010 185 of 186 DLM 74001 04 Rev A 186 ...
Page 186: ...Notes GE Analytical Instruments 2010 186 of 186 DLM 74001 04 Rev A 186 ...