DLM 74001-04 Rev. A
Chapter 5: Password Protection and DataGuard
To enable Password Protection
1. Press the
Menu
button and select the
DataGuard
tab.
2. Press the
Enable Password
button.
Password Protection security is immediately activated, and you will need to enter your Administrator ID
and Password the next time you access a Password-required screen. If the Lockout
On
option was
previously selected, the Main screen automatically displays after you press the
Enable Password
button. In this case, press
Menu
and enter your Administrator User ID and password.
To change the password
After Password protection has been enabled, you can change the default Password by following these steps:
1. Press the
Menu
button and select the
DataGuard
tab.
2. Press the
Change Password
button.
3. Enter the old password and press the
Enter
button.
4. Enter the new Password using the on-screen keyboard and press the
Enter
button.
5. Re-enter the same new Password for verification and press the
Enter
button.
The new Password is immediately activated.
To configure password settings
After enabling Password Protection, you can specify the password preferences as follows:
1. Press the
Menu
button and select the
DataGuard
tab.
2. Press the
Setup
button.
•
Password Expires
— Specifies how long the Password is valid before it must be changed. The
default is 90 days. You can enter a value in the range of 30 to 365 days.
•
Login Timeout
— Specifies the period of inactivity that passes before the use is automatically
logged out. The default is 5 minutes. You can enter a value in the range of 1 to 30 minutes.
•
Password Min Len
(minimum length)— Specifies the minimum number of characters required for a
valid Password. The default is 3 characters. You can enter a value in the range of 3 to 8 characters.
•
Menu Lockout
- When the
On
option is enabled, the user login and password are ONLY required to
access the Main menu. When the
Off
option is enabled, the user login and password are required to
change parameters that affect calibration or analysis measurements. The default setting is Menu
Lockout
Off
. For more information, see Figure 22: Menu Map — Menu Lockout Off (Requires
Calibration Passwords) on page 102.
Summary of Contents for Sievers 500 RL
Page 8: ...GE Analytical Instruments 2010 8 of 226 DLM 74001 04 Rev A ...
Page 10: ...GE Analytical Instruments 2010 10 of 226 DLM 74001 04 Rev A ...
Page 36: ...GE Analytical Instruments 2010 36 of 220 DLM 74001 04 Rev A ...
Page 66: ...GE Analytical Instruments 2010 66 of 226 DLM 74001 04 Rev A Chapter 3 Installation ...
Page 152: ...GE Analytical Instruments 2010 152 of 226 DLM 74001 04 Rev A Chapter 7 Maintenance ...
Page 170: ...GE Analytical Instruments 2010 170 of 226 DLM 74001 04 Rev A Chapter 8 Troubleshooting ...
Page 178: ...Appendix A GE Analytical Instruments 2010 178 of 186 DLM 74001 04 Rev A ...
Page 185: ...Notes GE Analytical Instruments 2010 185 of 186 DLM 74001 04 Rev A 186 ...
Page 186: ...Notes GE Analytical Instruments 2010 186 of 186 DLM 74001 04 Rev A 186 ...