Chapter 6: Calibration and System Protocols
If the label of the Conductivity Standard shows a value of 25 µS/cm HCl,
(or your configuration does
not measure for conductivity), press
Next
to continue.
If the label on the Conductivity Standard shows a different value,
press
Edit
, enter the value, press
Enter
, and then press
Next
to continue.
7. If you have a standard
iOS
System:
• Open the door to the
iOS
System and wait 30 seconds for water to drain.
• Insert the Verification Blank into the
iOS
System and press
Next
.
• When prompted, remove the Verification Blank from the
iOS
System, insert the Accuracy/Precision
and Verification Standard, and press
Next
.
• When prompted, remove the Accuracy/Precision and Verification Standard from the
iOS
System,
insert the Conductivity Standard, and press
Next
.
• When prompted, remove the Accuracy/Precision and Verification Standard (500 ppb sucrose) from
the
iOS
System. Confirm the value of the Conductivity Standard and insert the vial into the
iOS
System. If the label shows a value of 25 µS/cm HCl, press
Next
to continue. If the label shows a
different value, press
Edit
, enter the value, press
Enter
, and then press
Next
.
8. After the last standard has been analyzed, the summary screen is displayed. Part 1 of the summary
screen displays the data for the standards (see Figure 29). Press the
Next
button.
9. Press the
button if you want to print the Calibration results.
10. Part 2 of the summary screen calculations summary for the standards (see Figure 30). Acceptance
criteria are as follows:
• TOC Precision: RSD (Relative Standard Deviation
)
of last three measurements of 500 ppb standard
3% and standard deviation is > 2 ppb
• Conductivity Precision: RSD of last three compensated conductivity measurements on 25 µS/cm
standard
2%
• TOC Accuracy: % Difference
7%
• Conductivity Accuracy: % Difference
2%
The standard deviation and relative standard deviation are calculated as follows:
Standard Deviation =
n
x
2
x
2
–
n n
1
–
--------------------------------------
)
rejections
of
#
-
s
repetition
of
(#
set
a
in
ments
of Measure
Number
Result
Each
of
Sum
n
x
Summary of Contents for Sievers 500 RL
Page 8: ...GE Analytical Instruments 2010 8 of 226 DLM 74001 04 Rev A ...
Page 10: ...GE Analytical Instruments 2010 10 of 226 DLM 74001 04 Rev A ...
Page 36: ...GE Analytical Instruments 2010 36 of 220 DLM 74001 04 Rev A ...
Page 66: ...GE Analytical Instruments 2010 66 of 226 DLM 74001 04 Rev A Chapter 3 Installation ...
Page 152: ...GE Analytical Instruments 2010 152 of 226 DLM 74001 04 Rev A Chapter 7 Maintenance ...
Page 170: ...GE Analytical Instruments 2010 170 of 226 DLM 74001 04 Rev A Chapter 8 Troubleshooting ...
Page 178: ...Appendix A GE Analytical Instruments 2010 178 of 186 DLM 74001 04 Rev A ...
Page 185: ...Notes GE Analytical Instruments 2010 185 of 186 DLM 74001 04 Rev A 186 ...
Page 186: ...Notes GE Analytical Instruments 2010 186 of 186 DLM 74001 04 Rev A 186 ...