Chapter 6: Calibration and System Protocols
• (Optional, only if the pharmacopeia is set to include JP TOC
and
Conductivity) Remove the SDBS
standard, and insert the KCl standard into the
iOS
system, and press
Next
.
• When the measurements are completed, remove the vial from the
iOS
System, and slide the door
closed.
9. After the last standard has been analyzed, the JP Protocol screen appears. The Analyzer indicates
whether the JP TOC protocol passed
*
or failed based on the following criteria:
TOC Blank Average
250 ppb
Blank-Corrected Average
450 ppb
where
,
SDBS Average– TOC Blank Average = Blank-Corrected Average
The Analyzer will also report separately (
only
if JP Conductivity pharmacopeia is active)
whether the JP TOC and Conductivity Protocol passed or failed based on the following criteria:
TOC Blank Average
250 ppb
Blank-Corrected Average
450 ppb
Conductivity Difference
±
RSD
2%
Sample temperature is between15
⁰
C to 30
⁰
C
10. Complete the JP Protocol Worksheet with the pass or fail information.
11. Press the
Exit
button.
12. If you have a Super
iOS
system, and selected the
After
or the
Before and After
option, the Analyzer will
prompt you to remove the standards, and to insert the rinse cartridge or vials into the Super
iOS
system
to continue the Rinse activity. Take out the rinse cartridge or vials when completed.
13. If you have a standard
iOS
System, remove the standards and slide the
iOS
door closed.
*.
The Pass/Fail test uses rounded values.
Summary of Contents for Sievers 500 RL
Page 8: ...GE Analytical Instruments 2010 8 of 226 DLM 74001 04 Rev A ...
Page 10: ...GE Analytical Instruments 2010 10 of 226 DLM 74001 04 Rev A ...
Page 36: ...GE Analytical Instruments 2010 36 of 220 DLM 74001 04 Rev A ...
Page 66: ...GE Analytical Instruments 2010 66 of 226 DLM 74001 04 Rev A Chapter 3 Installation ...
Page 152: ...GE Analytical Instruments 2010 152 of 226 DLM 74001 04 Rev A Chapter 7 Maintenance ...
Page 170: ...GE Analytical Instruments 2010 170 of 226 DLM 74001 04 Rev A Chapter 8 Troubleshooting ...
Page 178: ...Appendix A GE Analytical Instruments 2010 178 of 186 DLM 74001 04 Rev A ...
Page 185: ...Notes GE Analytical Instruments 2010 185 of 186 DLM 74001 04 Rev A 186 ...
Page 186: ...Notes GE Analytical Instruments 2010 186 of 186 DLM 74001 04 Rev A 186 ...