GE Analytical Instruments ©2010
149 of 226
DLM 74001-04 Rev. A
Chapter 7: Maintenance
11. Reconnect the 1/4" Teflon tubing from the sampling port adapter to the inlet of the in-line filter. The in-
line filter has arrows on the body of the filter that indicate the direction of flow through the filter.
Tighten the nut one-quarter turn past finger-tight.
12. Reconnect the 1/4" Teflon tubing to the outlet of the filter and tighten the nut one-quarter turn past
finger-tight.
13. Turn on the water supply at the sampling port.
14. Turn on the Analyzer using the main power switch.
Figure 38: Replacing the In-Line Filter Element
Setting the Installation or Date for New Consumables
When you replace a consumable, you must enter the date of installation ensure that the Analyzer keeps an
accurate record of usage and that the indicators on the Main screen remain accurate.
1. Press the
Menu
button and select the
Maintenance
tab.
2. Press the
Consumables
button.
3. Press the
New...
button that corresponds to the consumable you are installing. Choose
New Lamp
,
New Tube
, or
New RBed
, press the
Accept
button to enter today’s date, or enter values for the
appropriate Day, Month and Year and then press the
Accept
button.
4. Repeat Step 3 for each consumable that you installed.
Note:
When Password protection or DataGuard is activated, not all users are able to modify
settings on this screen. See “Menu Map — DataGuard (Optional Upgrade)” on page 103 or
“Using DataGuard” on page 105 for more information.
60
M
Closed
End
Spring
Place wrenches
here
Loosen
inlet
adapter
60
M
Summary of Contents for Sievers 500 RL
Page 8: ...GE Analytical Instruments 2010 8 of 226 DLM 74001 04 Rev A ...
Page 10: ...GE Analytical Instruments 2010 10 of 226 DLM 74001 04 Rev A ...
Page 36: ...GE Analytical Instruments 2010 36 of 220 DLM 74001 04 Rev A ...
Page 66: ...GE Analytical Instruments 2010 66 of 226 DLM 74001 04 Rev A Chapter 3 Installation ...
Page 152: ...GE Analytical Instruments 2010 152 of 226 DLM 74001 04 Rev A Chapter 7 Maintenance ...
Page 170: ...GE Analytical Instruments 2010 170 of 226 DLM 74001 04 Rev A Chapter 8 Troubleshooting ...
Page 178: ...Appendix A GE Analytical Instruments 2010 178 of 186 DLM 74001 04 Rev A ...
Page 185: ...Notes GE Analytical Instruments 2010 185 of 186 DLM 74001 04 Rev A 186 ...
Page 186: ...Notes GE Analytical Instruments 2010 186 of 186 DLM 74001 04 Rev A 186 ...