Chapter 3: Installation
which is adjusted by the screw on the sample inlet system. Turn the screw clockwise to decrease flow,
and turn the screw counter-clockwise to increase flow.
Step 7: Fill the DI Water Cartridge
Fill the DI water cartridge with DI water, using the squeeze bottle provided in the accessories kit.
1. Fill the water bottle with DI water.
2. Open the Analyzer’s front panel (you will need the key from the accessories kit to open the bottom
latch).
3. Locate the DI water cartridge at the center of the Analyzer (see Figure 7)
Figure 7: Locating The DI Water Cartridge
4. Lift the spring clip (see Figure 8 on page 57) at the top of the DI water cartridge, and rotate the cartridge
clockwise (as viewed from the top) so it slides out of the retaining hooks.
Note:
Setting a water flow rate that exceeds 300 mL/min may result in poor system
suitability verification data or any measurements taken from sample water in vials.
Excessive flow also will not allow the sample pump to draw sample properly.
Note:
The
iOS
System door may be taped closed for safety during transport. Remove the
tape before proceeding.
Summary of Contents for Sievers 500 RL
Page 8: ...GE Analytical Instruments 2010 8 of 226 DLM 74001 04 Rev A ...
Page 10: ...GE Analytical Instruments 2010 10 of 226 DLM 74001 04 Rev A ...
Page 36: ...GE Analytical Instruments 2010 36 of 220 DLM 74001 04 Rev A ...
Page 66: ...GE Analytical Instruments 2010 66 of 226 DLM 74001 04 Rev A Chapter 3 Installation ...
Page 152: ...GE Analytical Instruments 2010 152 of 226 DLM 74001 04 Rev A Chapter 7 Maintenance ...
Page 170: ...GE Analytical Instruments 2010 170 of 226 DLM 74001 04 Rev A Chapter 8 Troubleshooting ...
Page 178: ...Appendix A GE Analytical Instruments 2010 178 of 186 DLM 74001 04 Rev A ...
Page 185: ...Notes GE Analytical Instruments 2010 185 of 186 DLM 74001 04 Rev A 186 ...
Page 186: ...Notes GE Analytical Instruments 2010 186 of 186 DLM 74001 04 Rev A 186 ...