Section 8-3 - System Software - Installation/Upgrade Procedure
8-3-5-2-1-8
Connectivity - Recording other settings for Print
1) From the Touch Panel, select
Utility -> Connectivity -> Buttons
.
Figure 8-8 Button settings
8-3-5-2-1-9
Connectivity - Service settings
1) From the touch panel, select Utility -> Connectivity -> Service.
Table 8-12 Record settings from Button screen
Still Images
Format
Ultrasound Image / Secondary Capture - Image / Secondary Capture - Video /
Secondary Capture - Screen
Compression
None / RLE / JPEG / JPEG2000
Clip/Volumes
Clips Add Multiframe Data
ON / OFF
Compression
None / RLE / JPEG / JPEG2000
Quality
Volume File Format
1 - Standard DICOM / 2 - Standard DICOM with Raw / 3 - Enhanced DICOM / 2&3 (2 files)