Section 5-11 - Common Service Desktop (R3 only)
To run a diagnostic
:
1.) Navigate to
Utility (second page) > Service > Utilities > Disruptive Mode Utility
and set
Disruptive mode to On.
2.) Navigate to
Utility (second page) > Service > Diags > Run Diags
.
3.) From the
Service Menu
tab, select the diagnostic (for example, GRLY with Probes). The
Selected
Diagnostics
and
Diagnostic Detail
areas populate with information specific to the selected
diagnostic.
4.) In
Execution Mode
, select the desired mode.
5.) In
Loop Count
, select the number of times to run the diagnostic.
6.) Click
Start
.
7.) View progress in the
Total Progress
area.
8.) View the results (Passed or Failed) in the
Diagnostic Results Summary
area.
9.) To view details about the diagnostics run, see
.