Chapter 8 - Replacement Procedures
8-183
8-46-5
Functional Check
1) Turn on the circuit breaker.
2) Boot-up the console using power button on OPIO ASSY.
3) Check that battery icon is displayed on left bottom side.
4) Switch off the circuit breaker (Wall power mode -> Battery mode). Confirm that is operating with
battery mode.
Figure 8-222 Turning on circuit breaker
Figure 8-223 Booting-up using power button
Figure 8-224 Checking battery icon