Chapter 8 - Replacement Procedures
8-109
8-17-10
Separating MDC assy from MCB assy
1) Remove 7 screws around MDC assy. Refer to following figure.
Figure 8-120 Screw points around MDC assy
2) Carefully separate MDC assy from MCB assy. Refer to following figure.
Figure 8-121 Removal of MDC assy