Chapter 5 - Components and Functions (Theory)
5-47
This table shows all the elements available on
System Status
with descriptions.
For more information, see
:
•
•
•
•
•
•
•
Table 5-16 System Status
Element
DESCRIPTION
Hardware Configuration
Displays the contents of the hardware configuration file.
Temperature
Displays the temperature monitoring of the LOGIQ P9/P7 as normal, warning, or
error using the standard colors and icons. Includes temperature out-of-spec events
including an overall count of event warnings and errors for the last 30 days or the
last status checkpoint whichever is sooner. Temperature displays temperature a
graph for the configured monitoring points.
Voltage
Displays the voltage monitoring of the LOGIQ P9/P7 as normal, warning or error
using the standard colors and icons. Includes voltage out-of-spec events including
an overall count of the event warnings and errors for the last 30 days or last status
checkpoint whichever is sooner.
Bootup Issues
Based on the latest occurrence, displays all abnormal boots of the LOGIQ P9/P7 for
either the past 30 days or the latest check point.
Alert Logs
Based on the latest occurrence, displays all alerts generated by the LOGIQ P9/P7
for either the past 30 days or the latest check point.
USB Devices
Displays the USB connected devices and their status.
System On Time
Displays the time the LOGIQ P9/P7 was turned on.
Time Since Last Reboot
Displays the time the LOGIQ P9/P7 was last rebooted.