Chapter 8 - Replacement Procedures
8-107
8-17-8
Removal Procedure - Separating MPI and MCB
1) Unscrew 6 screws. Refer to following figure.
Figure 8-116 Screw points on MPI assy
2) Carefully separate MPI assy from MCB assy. Refer to following figure.
Figure 8-117 Lifting up MPI and position of Stacking up connector on MPI assy