Chapter 3 - Setup Instructions
3-11
Section 3-6
Connection of Auxiliary Devices
NOTE:
Normally auxiliary devices and peripherals come pre-installed with the system.
below outlines hardware installation procedures described in the sub-sections.
NOTE:
For more detailed Safety Considerations when connecting auxiliary devices to the LOGIQ P9/P7
system, please review:
Chapter 1 - Auxiliary Devices Safety, on page 1-10
Table 3-7
Connection Procedures
Sub-section
Description
Page Number
Connecting the Black & White Printer
3-6-3
Connecting the Color Printer
3-14
Connecting the Secondary “Patient” LCD Monitor
Connecting the USB Flash Memory Stick
Connecting the external USB Hard disk (Handydrive)
General Remarks and Hints when using external USB-Devices
WARNING
WARNING
!! WARNING:
After each installation, the leakage currents have to be measured according to
IEC 60601-1 respectively ANSI/AAMI ES60601-1.
CAUTION
!! CAUTION:
Please observe that some printers may not be medical devices! If the Bluetooth Printer and/or
Line Printers are not medical devices, they have to be located outside of the patient
environment (according to IEC 60601-1 / ANSI/AAMI ES60601-1).
2,5 m
1,5 m
1,5 m
1,5 m