GE
RAFT
LOGIQ P9/P7
D
IRECTION
5604324
, R
EVISION
11
DRAFT (J
ANUARY
24, 2019)
S
ERVICE
M
ANUAL
8-172
Section 8-43 - Replacement of DCWD ASSY
Section 8-43
Replacement of DCWD ASSY
8-43-1
Manpower
1 person, 15 minutes
8-43-2
Tools
Standard Phillips screwdriver.
8-43-3
Removal Procedure
1) Separate MPI assy from MCB and MDC. Refer to the
Removal Procedure -
Separating MPI and MCB.
2) Remove 3 screws on MPI assy.
Figure 8-203 DCWD Board screw points
3) Carefully separate DCWD board from MPI assy not to damage the DCWD connector. Refer to
following figure.
Figure 8-204 Position of DCWD board connector