Chapter 5 - Components and Functions (Theory)
5-65
This table shows all the elements available on
Run Diags
with descriptions.
Table 5-26 Run Diags
Element
DESCRIPTION
Service Menu
Available service diagnostics in a diagnostics tree.
Engineering Menu
Available engineering diagnostics in a diagnostics tree.
Clear All
Resets the selections within the diagnostics tree.
Selected Diagnostics
Diagnostics that have been selected.
Diagnostic Detail
Test Name
Name of the diagnostic.
Last Execution Result
Result of the diagnostic run.
Expected Execution Time
Time expected to run the diagnostic.
Maximum Execution Time
Maximum amount of time allowed to run the diagnostic.
Instruction
Special instructions needed in order to perform the diagnostic.
Diagnostic Controls
Execution Mode
Way the diagnostic behaves when a failure occurs. Valid values are:
• Continue on Failure
• Finish Loop and Stop on Failure
• Stop on Failure
Loop Count
Number of times to run the selected diagnostic.
Start
Starts the selected diagnostic.
Stop
Stops the selected diagnostic.
Reset
Clears the current diagnostic information.
Total Progress
Current Progress
Progress the current diagnostic.
Total Progress
Progress for all of the selected diagnostics.
Diagnostic Results Summary
Diag Name
Name of the diagnostic.
Date/Time
Date and time of the diagnostic run.
Diag Run Time
Amount of time taken to run the diagnostic.
Overall Status
Current status of the diagnostic.
Pass/Fail
Whether the selected diagnostic passed or failed.
• Green - Diagnostic passed
• Red - Diagnostic failed