Revision C
Dash 2000 Patient Monitor
4-13
2005873-001
If you want to move a call to another level, follow this procedure. (In this
example we will change V TACH from a Crisis level to a Warning level.)
Any changes made with this menu option are temporary and will revert to
defaults upon discharge.
NOTE: Asystole and ventricular fibrillation (V FIB, VTAC)
cannot be
moved when in ICU (Adult and Neonatal) Mode.
1. Rotate the Trim Knob control to move the pointer in front of V TACH.
2. Press the Trim Knob control. The level for the V TACH call highlights.
3. Rotate the Trim Knob control until WARNING is displayed.
4. Press the Trim Knob control to complete the change. The information
window is reorganized to include V TACH as a WARNING alarm and
the change is in effect.
The information window remains open for you to make any other changes.
5. When you have completed all changes, move the pointer to RETURN
and press the Trim Knob control to close the information window.
Summary of Contents for marquette Dash 2000
Page 1: ...Dash 2000 Patient Monitor Software Version 3 Operator s Manual 2005873 001 ENG Revision C ...
Page 64: ...2 14 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Page 82: ...3 18 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Page 96: ...4 14 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Page 122: ...5 26 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Page 130: ...6 8 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Page 140: ...7 10 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Page 168: ...8 28 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Page 184: ...9 16 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Page 202: ......