8-12
Dash 2000 Patient Monitor
Revision C
2005873-001
Smart-Lead Fail
Leads Fail Patient
Condition
NOTE: The smart-lead fail feature works only when using a 5-leadwire
electrode configuration.
The smart-lead fail feature is continually checking the integrity of the electrodes
to allow uninterrupted monitoring. Should the quality of an electrode signal
degrade to an inadequate level, a lead fail message will be displayed. If the
lead fail affects the ECG waveform monitored in the top trace position on the
screen, monitoring will
automatically switch to another lead. Refer to the
chart below.
Message
IEC Label
New lead monitored
RA FAIL
R
Lead III
LL FAIL
F
Lead I
LA FAIL
L
Lead II
V FAIL
C
Lead II
LEADS FAIL
No waveform displayed—right leg
fail or multiple leads fail
Individual lead fail messages will not be displayed when using single-lead
analysis, but lead switching will occur.
There is a system alarm to alert you when more than one lead fails. In OR
mode, the monitor defaults this ”leads fail patient” condition as a system
ADVISORY alarm. You can, however, set it as a CRISIS alarm in Monitor
Defaults.
In Adult-ICU and Neonatal-ICU modes this alarm defaults to a system
WARNING alarm but can also be set to a CRISIS alarm. Refer to Setup Default
Display section in Chapter 5, Monitor Setup.
Summary of Contents for marquette Dash 2000
Page 1: ...Dash 2000 Patient Monitor Software Version 3 Operator s Manual 2005873 001 ENG Revision C ...
Page 64: ...2 14 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Page 82: ...3 18 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Page 96: ...4 14 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Page 122: ...5 26 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Page 130: ...6 8 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Page 140: ...7 10 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Page 168: ...8 28 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Page 184: ...9 16 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Page 202: ......
Page 220: ...11 18 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...