Revision C
Dash 2000 Patient Monitor
3-15
2005873-001
Set Graph Location
ECG Source
The GRAPH LOCATION option display a menu which allows you to choose
writers for alarm and manual graphs. This option is used mostly if the
Dash does not have the optional built-in writer. The graph location should
be changed whenever the Dash is moved from one unit to another. Refer to
GRAPH LOCATION in the Monitor Setup chapter for more detailed
information.
This menu option is found in Combo and Rover Combo admit menus.
This menu option, which is unique to those monitors which can obtain
parameter data from telemetry or the monitor, is used to tell the monitor
where to get the parameter information.
Follow this procedure:
1. Select ECG SOURCE option from the Admit Menu. A popup menu and
an information window are displayed.
21-NOV-2000 16:27
II
DAK.BED 1
SELECT ECG SOURCE
MONITOR
331
446
447
783
MAIN
MENU
ECG
SOURCE
PREV.
MENU
>
040
ADMIT
PATIENT
SET
GRAPH
LOCATION
ECG SOURCE Popup Menu and Information Window
The information window lists the possible sources from which the
parameter data can be obtained. The transmitter numbers listed are
those transmitters assigned to the unit.
2. Rotate the Trim Knob control to move the pointer (>) in front of the
source of choice. Select MONITOR if your patient’s ECG cable is
connected to the monitor. If your patient is wearing a telemetry
transmitter, select the transmitter number that matches the one on the
back of the transmitter.
3. Press the Trim Knob control to close the popup menu and information
window. If you choose a transmitter number, that number (TTX) needs
to be admitted before it will appear in the ECG parameter window.
4. Select ADMIT PATIENT to admit the transmitter. The TTX number
now appears in the ECG parameter window.
Summary of Contents for marquette Dash 2000
Page 1: ...Dash 2000 Patient Monitor Software Version 3 Operator s Manual 2005873 001 ENG Revision C ...
Page 64: ...2 14 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Page 82: ...3 18 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Page 96: ...4 14 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Page 122: ...5 26 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Page 130: ...6 8 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Page 140: ...7 10 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Page 168: ...8 28 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Page 184: ...9 16 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Page 202: ......
Page 220: ...11 18 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...