Revision C
Dash 2000 Patient Monitor
8-25
2005873-001
Troubleshooting
Problem:
Inaccurate heart rate
and/or false asystole
Solution:
Check ECG signal from patient:
1. Check/adjust lead placement.
2. Check/perform skin preparation.
3. Check/replace
electrodes.
Check amplitude of ECG waveform:
1. Select ECG parameter label.
2. Select DISPLAY LEAD.
3. Scroll through all ECG leads and check for 0.5 mV amplitude at
normal (1X) size. (At least 0.5 mV amplitude is required for QRS
detection.) For borderline signals, validate on a graph.
4. If amplitudes are low, electrodes may need to be repositioned or
replaced.
Relearn arrhythmia:
1. Select ECG parameter label.
2. Select
RELEARN.
IF PROBLEM CONTINUES AND ECG AMPLITUDE IS LOW:
1. Select ECG parameter label.
2. Select ECG SIZE.
3. Select 2X or higher.
4. Close the popup menu.
5. Select
RELEARN.
IF PROBLEM CONTINUES: Change to single lead ECG detection and
processing:
1. Select ECG parameter label.
2. Select ANALYSIS SETTINGS.
3. Select LEAD ANALYSIS.
4. Select SINGLE LEAD.
5. Go to PREV MENU.
6. Select DISPLAY LEAD.
7. Assess all the ECG leads and change top ECG waveform to display
lead with greatest amplitude. (At least 0.5 mV amplitude is
required for QRS detection.)
Summary of Contents for marquette Dash 2000
Page 1: ...Dash 2000 Patient Monitor Software Version 3 Operator s Manual 2005873 001 ENG Revision C ...
Page 64: ...2 14 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Page 82: ...3 18 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Page 96: ...4 14 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Page 122: ...5 26 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Page 130: ...6 8 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Page 140: ...7 10 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Page 168: ...8 28 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Page 184: ...9 16 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Page 202: ......
Page 220: ...11 18 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...