9-14
Dash 2000 Patient Monitor
Revision C
2005873-001
Disconnect Alarm
The DISCONN ALARM feature is found in ART and FEM Settings menus.
NOTE:
This feature is not available when the monitor is set up for
Neonatal-ICU mode.
If the mean pressure falls below 25 mmHg and the disconnect alarm is on, a
Warning alarm sounds and the message DISCONNECTED is displayed in the
values window. Check your patient immediately in the event the catheter has
dislodged.
To turn this feature off and on, select DISCONN ALARM option from the
appropriate Settings menu. This feature can be set off or on in Monitor
Defaults. You cannot change the alarm level. It will always be a warning
alarm.