10-12
Dash 2000 Patient Monitor
Revision C
2005873-001
NBP Limits
This option provides an information window that shows the alarm limits for
systolic, diastolic, and mean pressures. A new set of menu options is
displayed to allow you to adjust these limits.
Follow this procedure:
1. Select NBP LIMITS from the NBP Menu. An information window is
displayed on the screen and a new set of menu options is displayed in
the menu area.
21-NOV-2000 16:27
II
DAK.BED 1
NBP LIMITS
RETURN
NBP-S
NBP-D
NBP-M
UNITS
mmHg
mmHg
mmHg
LOW
20
20
40
HIGH
200
120
140
MAIN
MENU
NBP
LIMITS
>
112
NBP LIMITS Menu and Information Window
2. Rotate the Trim Knob control to move the pointer (>) through the list.
3. Press the Trim Knob control when the pointer is in front of the
parameter for which you want to change default limits. The parameter
label highlights.
4. Turn the Trim Knob control to highlight the low or high limit.
5. Press the Trim Knob control again and rotate to change value. Press
the Trim Knob control to complete the change.
6. Rotate the Trim Knob control to highlight the parameter label and
press to unhighlight.
7. Select RETURN when all changes are completed. The popup menu and
information window close.
Summary of Contents for marquette Dash 2000
Page 1: ...Dash 2000 Patient Monitor Software Version 3 Operator s Manual 2005873 001 ENG Revision C ...
Page 64: ...2 14 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Page 82: ...3 18 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Page 96: ...4 14 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Page 122: ...5 26 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Page 130: ...6 8 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Page 140: ...7 10 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Page 168: ...8 28 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Page 184: ...9 16 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Page 202: ......
Page 220: ...11 18 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...