5-20
Dash 2000 Patient Monitor
Revision C
2005873-001
Alarm Graph
Speed
This menu option allows you to turn automatic alarm graphs off and on.
This is a direct action menu option. Selecting this option switches back and
forth between off and on.
MAIN
MENU
ECG 1:
LEAD III
WAVEFORM
2:
ART
WAVEFORM
3:
OFF
ALARM
GRAPH:
ON
GRAPH
TIME /
SPEED
PREV.
MENU
GRAPH
LOCATION
060
ALARM GRAPH Option
An arrhythmia alarm graph continues to run until a normal sinus rhythm
is again detected, or until you manually stop it at the writer. All other
alarm graphs will run for 20-seconds.
NOTE: To convert an arrhythmia alarm graph to a manual graph, you
must stop the graph using the GRAPH GO / STOP key on the
writer and immediately restart the graph.
The speed at which a graph strip is printed is controlled with this menu
option. The slower the speed, the more condensed the data will be. This
option affects manual graphs only and must be selected before starting the
graph.
NOTE: Alarm graphs always run at 25 millimeters per second.
1. Select the GRAPH TIME/SPEED option from the Graph Setup Menu.
A menu with the two options SPEED and TIMED GRAPH appears.
2. Select the SPEED option. A popup menu opens displaying all choices.
MAIN
MENU
SPEED:
25
0.1
0.5
1
5
PREV.
MENU
10
12.5
>25
50
061
SPEED Popup Menu
NOTE: A laser printer does not support the 0.1 millimeters per second
speed.
3. Rotate the Trim Knob control so the pointer is in front of the speed you
want.
4. Press the Trim Knob control to confirm the change and close the popup
menu.
Summary of Contents for marquette Dash 2000
Page 1: ...Dash 2000 Patient Monitor Software Version 3 Operator s Manual 2005873 001 ENG Revision C ...
Page 64: ...2 14 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Page 82: ...3 18 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Page 96: ...4 14 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Page 122: ...5 26 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Page 130: ...6 8 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Page 140: ...7 10 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Page 168: ...8 28 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Page 184: ...9 16 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Page 202: ......
Page 220: ...11 18 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...