12-10
Dash 2000 Patient Monitor
Revision C
2005873-001
Cardiac Artifact
Alarm
The cardiac artifact alarm alerts you to the fact that the respiration rate is within
5% of the heart rate (over 30 consecutive breaths). If this happens, the
respiration program may be counting heart beat artifact as respiration. The
cardiac artifact alarm is an Advisory alarm—ARTIFACT is displayed in the RR
values window, and a 1-beep tone sounds.
There is no adjustable limit for this alarm, but you can turn it off and on.
WARNING
If the cardiac artifact alarm is turned off, apnea events may not
be detected.
Select CARDIAC ARTIFACT option from the Respiration Menu to turn the
alarm off and on. This is a direct action menu option.