Revision C
Dash 2000 Patient Monitor
10-5
2005873-001
Checklist
1. A cuff appropriate for limb size has been selected.
2. Cuff is properly placed on patient.
3. Patient cable is connected to monitor.
NOTE: Do not connect more than one NBP patient cable.
4. Tubes between the cuff and the monitor are not kinked or blocked.
5. Correct CUFF SIZE has been selected from the NBP menu.
6. Confirm that the cuff is fully deflated before positioning on the patient.
7. Start an NBP reading following auto or stat mode procedures as
detailed in this section, or use the NBP GO/STOP control key.
8. The measurement should be performed on the left arm if feasible. Roll
up sleeve before measurement. Only very thin fabrics will not impair
the measurement.
9. Check that rolled up sleeve does not constrict the vessels.
10. The arm should be placed on a surface that is level with the heart. The
palm of the hand should face up.
11. Position patient so that no external pressure is applied against cuff
while monitoring. External pressure may cause inaccurate blood
pressure values.
WARNING
The cuff size selected in the NBP menu and the cuff size used
must be correct to obtain reliable NBP data and to prevent
overpressure in neonatal or pediatric use.
Summary of Contents for marquette Dash 2000
Page 1: ...Dash 2000 Patient Monitor Software Version 3 Operator s Manual 2005873 001 ENG Revision C ...
Page 64: ...2 14 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Page 82: ...3 18 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Page 96: ...4 14 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Page 122: ...5 26 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Page 130: ...6 8 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Page 140: ...7 10 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Page 168: ...8 28 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Page 184: ...9 16 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Page 202: ......
Page 220: ...11 18 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...