10-8
Dash 2000 Patient Monitor
Revision C
2005873-001
Systolic Search
NBP Go/Stop Key
Power Key
NOTE: The cuff target pressure must be higher than the patient's systolic
pressure to obtain an accurate systolic and diastolic reading.
If a systolic blood pressure is not found, the Dash monitor will search for a
systolic value at a cuff pressure higher than the initial target pressure. This will
occur only once per NBP determination. The maximum pressure allowed in
systolic search is limited by the normal range for cuff pressures.
The NBP GO/STOP key on the front of the monitor is a quick way to take one
measurement
without going into the NBP menu. It can also be used to stop
a measurement already in progress.
If you turn the power off with the POWER key on the front of the monitor,
NBP is also turned off. This applies to both auto NBP (if running) and
manual NBP. Turning the power on again enables manual NBP. It does not
automatically restart auto NBP.
Summary of Contents for marquette Dash 2000
Page 1: ...Dash 2000 Patient Monitor Software Version 3 Operator s Manual 2005873 001 ENG Revision C ...
Page 64: ...2 14 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Page 82: ...3 18 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Page 96: ...4 14 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Page 122: ...5 26 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Page 130: ...6 8 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Page 140: ...7 10 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Page 168: ...8 28 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Page 184: ...9 16 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Page 202: ......
Page 220: ...11 18 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...