10-16
Dash 2000 Patient Monitor
Revision C
2005873-001
Troubleshooting
NBP Status
Messages
A status message will be displayed in the NBP values window if a
measurement is unable to be completed. Following is an alphabetical list of the
status messages with monitor response and action to take.
STATUS MESSAGE
MONITOR RESPONSE
SOLUTION
CUFF INFLAT OVER 5 MINS
System status alarm. Auto mode
will shut off.
Check cuff, if no air is in cuff, try
another measurement. If problem
persists, contact service.
If air is in cuff, remove cuff and
contact service.
DEFLATION FAIL
System status alarm. Auto mode
will shut off.
Remove cuff and contact service.
HARDWARE FAIL
System status alarm. Auto mode
will shut off.
Contact service.
NO DETERMINATION
System status alarm. Auto mode
will shut off.
Check patient and cuff placement;
try another measurement. If
problem persists, contact service.
OVER PRESSURE
System status alarm.
Remove cuff and contact service.
PUMP TIMEOUT
(inflation failure/pressure leak)
System status alarm.
Check connections between cuff
and monitor; try another
measurement. If problem persists,
contact service.
Check for worn connector ”O”
rings. Contact service.
TOTAL TIMEOUT
(measurement > 3 minutes)
System status alarm. Auto mode
will shut off.
Possible excessive patient
movement or arrhythmia
condition. Check patient; try
another measurement.
For neonatal mode, the status
message will appear after 90
seconds.
Any two consecutive NBP status messages will shut Auto mode off if it is
running.
A message will clear when the next measurement is initiated, or a message
can be cleared manually with the CLEAR NBP READING option in the NBP
menu.
An NBP status message will also be included in the graph header in an
abbreviated form, when applicable.
Summary of Contents for marquette Dash 2000
Page 1: ...Dash 2000 Patient Monitor Software Version 3 Operator s Manual 2005873 001 ENG Revision C ...
Page 64: ...2 14 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Page 82: ...3 18 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Page 96: ...4 14 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Page 122: ...5 26 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Page 130: ...6 8 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Page 140: ...7 10 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Page 168: ...8 28 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Page 184: ...9 16 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Page 202: ......
Page 220: ...11 18 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...