Revision C
Dash 2000 Patient Monitor
10-11
2005873-001
NBP Stat
Early Systolic
Measurement
NOTE: The NBP STAT feature is not available when the monitor is set up
for Neonatal-ICU mode. Please do not use the STAT mode on
neonatal patients when the monitor is set up for Adult-ICU or
Operating Room mode.
NBP STAT enables 5 minutes of continuous, sequential, automatic NBP
measurements.
Follow this procedure:
•
Activate the NBP STAT option of the NBP menu to start 5 minutes of
continuous, sequential NBP measurement.
•
Press OFF to discontinue the STAT process before the 5-minute period
is up, or simply press NBP GO / STOP key on the front of the monitor.
Early systolic measurement is a feature of the NBP STAT mode. As soon as
you enter NBP STAT mode, cuff inflation begins, a measurement is taken,
and the systolic, diastolic, and mean values are displayed. Thereafter, the
systolic value will be displayed shortly after the measurement starts. When
the measurement is complete the monitor will beep and the diastolic and
mean values will be displayed.
Summary of Contents for marquette Dash 2000
Page 1: ...Dash 2000 Patient Monitor Software Version 3 Operator s Manual 2005873 001 ENG Revision C ...
Page 64: ...2 14 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Page 82: ...3 18 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Page 96: ...4 14 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Page 122: ...5 26 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Page 130: ...6 8 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Page 140: ...7 10 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Page 168: ...8 28 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Page 184: ...9 16 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Page 202: ......
Page 220: ...11 18 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...