Guidelines When
Doing Combination
Monitoring
For Which
Application is the
Monitor Set?
Here are some guidelines to remember when doing combination (Combo
and Rover Combo) monitoring:
•
When monitoring ECG from telemetry, ECG limits and arrhythmia
alarm levels are not your monitor defaults but are the telemetry
defaults or as modified at the central station. You can, however, adjust
these settings at the monitor.
•
When ECG monitoring is switched from telemetry to the monitor,
telemetry is automatically discharged.
•
When monitoring ECG from telemetry, the alarm pause feature (if
available on your transmitter) is honored at the monitor.
•
If you switch ECG monitoring from telemetry to monitor, the ECG
limits, arrhythmia alarm levels, and display formats will be recalled
from Monitor Defaults.
•
The arrhythmia alarm level for the telemetry will be reduced to match
the monitor's arrhythmia alarm level.
If you want to know the application for which the monitor is set, follow
these steps:
1. Select MORE MENUS option from the Main Menu.
2. Select MONITOR SETUP.
3. Select SOFTWARE SUMMARY. An information window is displayed.
(See the figure below.)
21-NOV-2000 16:27
II
DAK.BED 1
SOFTWARE SUMMARY
SOFTWARE REVISIONS:
MAIN
MAIN BOOT
DAS
DAS BOOT
DDW
ADMIT MENU:
MONITOR TYPE:
3AX01
2ATST
4A
2BX01
VER 2A
STANDARD
ADLT-ICU
MAIN
MENU
MONITOR
DEFAULTS
PREV.
MENU
SOFTWARE
SUMMARY
OK
PARAM.
ON / OFF
>
070
(C) COPYRIGHT 2000 - 2001
GE MEDICAL SYSTEMS
INFORMATION TECHNOLOGIES GMBH
SOFTWARE SUMMARY Information Window
Summary of Contents for marquette Dash 2000
Page 1: ...Dash 2000 Patient Monitor Software Version 3 Operator s Manual 2005873 001 ENG Revision C ...
Page 64: ...2 14 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Page 82: ...3 18 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Page 96: ...4 14 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Page 122: ...5 26 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Page 130: ...6 8 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Page 140: ...7 10 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Page 168: ...8 28 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Page 184: ...9 16 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Page 202: ......
Page 220: ...11 18 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...