14-2
Dash 2000 Patient Monitor
Revision C
2005873-001
Defib Sync/Analog Output
ECG Acquisition
Modules
The configuration for the analog output from the ECG acquisition modules
is detailed below. Analog output is obtained through the Defib Sync
connector on the rear of the monitor.
ECG ANALOG OUTPUT:
Top displayed lead including augmented leads. If pace is on there is an
enhanced pacer pulse (5 volts at 2 milliseconds) when pacemaker triggering
is detected.
In the event of a LEAD FAIL, refer to the following:
Lead Failed
AHA Label
IEC Label
ECG Analog Out
Right arm
RL
R
III
Left arm
LA L
II
Left leg
LL
F
I
Chest V C
II
Right leg
RL
N
Buzz*
*A right leg lead fail causes an all LEADS FAIL on the screen and a buzz is
output.
NOTES:
1. All ECG outputs are 1 volt per millivolt referred to input.
2. In single lead mode, if the specified lead is not available, a buzz will be
output.
DIGITAL SYNCHRONIZATION OUTPUT:
A digital sync output is available with a return marker. For the technical
specifications refer to appendix Technical Specifications, section Analog
Out and Defibrillator Synchronization Pulse.
Summary of Contents for marquette Dash 2000
Page 1: ...Dash 2000 Patient Monitor Software Version 3 Operator s Manual 2005873 001 ENG Revision C ...
Page 64: ...2 14 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Page 82: ...3 18 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Page 96: ...4 14 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Page 122: ...5 26 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Page 130: ...6 8 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Page 140: ...7 10 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Page 168: ...8 28 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Page 184: ...9 16 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Page 202: ......
Page 220: ...11 18 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...