12-12
Dash 2000 Patient Monitor
Revision C
2005873-001
Troubleshooting
Respiratory
Waveform
Cardiac Artifact
Illustrated below is a respiratory waveform which is regular and even. The
inspiration and expiration markers are identified.
Inspiration Marker
130
Expiration Marker
Regular and Even
In cases where there is cardiac artifact, if the detection sensitivity percentage is
set too low, the artifact will be detected as breaths.
Increase the detection
sensitivity percentage until the markers correctly identify each inspiration
and expiration. See the markers in the figures below. (A = artifact, B =
breath)
A
B
A
A
B
131
Incorrect Detection
Correct Detection