Revision C
Dash 2000 Patient Monitor
1-3
2005873-001
On the back of the monitor, you can find all connectors for equipment and
network. (See the illustration below.)
NOTE: Refer to the service manual for system safety requirements when
connecting the monitor to accessory equipment.
Aux Port – Used for
nurse call connector
only.
Defib Sync Connector – Provides ECG analog
output signals to user-supplied equipment. A 5-volt,
2-milisecond artificial pacer spike is added to the
analog output when PACE is on and detection
occurs. Refer to the Appendices, Analog Output, for
details on signal output.
Network Connector – A cable
can be connected to this port
for monitors used in the patient
monitoring network configuration.
002
Handle – An optional alarm indicator can be
built into the handle of the monitor (not shown).
When activated, the LED indicator flashes red
for CRISIS and WARNING patient status alarms
and yellow for all other alarms.
AC Power
Connector
Peripheral
Expansion
Connector
Monitor, Back View
Optional Nurse Call
For connection of the monitor to a nurse-call system, a separation line
according to IEC 60601 (4 kV voltage isolation) is required. Use parts set
2005693-001. Please note the information given in chapter 4, section
"Alarm Structure", on the different categories of transferred alarms.
Summary of Contents for marquette Dash 2000
Page 1: ...Dash 2000 Patient Monitor Software Version 3 Operator s Manual 2005873 001 ENG Revision C ...
Page 64: ...2 14 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Page 82: ...3 18 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Page 96: ...4 14 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Page 122: ...5 26 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Page 130: ...6 8 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Page 140: ...7 10 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Page 168: ...8 28 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Page 184: ...9 16 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Page 202: ......
Page 220: ...11 18 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...