Dash 2000 Patient Monitor
9-15
2005873-001
Troubleshooting
Problem:
Displayed pressure values
are different than expected
Problem:
Smart BP is on. Artifact is
sensed without flush, draw,
or zero
Solution
:
Check tubing for bubbles.
Remove excessive tubing.
Check phlebostatic axis placement of transducer.
Rezero pressure:
1. Place the transducer at the left atrial level.
2. Close transducer stopcock to patient.
3. Open the venting stopcock to air.
4. Press FUNCTION key on the front panel of the monitor.
5. Close the venting stopcock to air.
6. Open transducer stopcock to patient.
Solution:
Turn Smart BP OFF then ON. If problem persists, you may need to disable
Smart BP by turning it off for that pressure.
Use Alarm Pause feature prior to drawing blood to reduce unnecessary alarms
if Smart BP is disabled.
Summary of Contents for marquette Dash 2000
Page 1: ...Dash 2000 Patient Monitor Software Version 3 Operator s Manual 2005873 001 ENG Revision C ...
Page 64: ...2 14 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Page 82: ...3 18 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Page 96: ...4 14 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Page 122: ...5 26 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Page 130: ...6 8 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Page 140: ...7 10 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Page 168: ...8 28 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Page 184: ...9 16 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Page 202: ......
Page 220: ...11 18 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...