Revision C
Dash 2000 Patient Monitor
5-7
2005873-001
Setup Default Limits
You can set default limits for all parameters also.
1. Select DEFAULT LIMITS from the Monitor Defaults Menu. A popup
menu and information window are displayed.
21-NOV-2000 16:27
II
DAK.BED 1
SETUP DEFAULT LIMITS
RETURN
HR
SPO2
RR
RR-APNEA
NBP-S
NBP-D
NBP-M
TP
UNITS
BPM
%
#/MIN
SEC
mmHg
mmHg
mmHg
˚C
LOW
50
90
5
80
20
40
30.0
HIGH
150
105
55
20
200
120
140
42.0
MAIN
MENU
DEFAULT
LIMITS
PREV.
MENU
>
066A
DEFAULT LIMITS Popup Menu and Information Window
2. Rotate the Trim Knob control to move the pointer (>) through the list.
3. Press the Trim Knob control when the pointer is in front of the
parameter for which you want to change default limits. The parameter
label highlights.
4. Turn the Trim Knob control to highlight the low or high limit.
5. Press the Trim Knob control again and rotate to change the value.
Press the Trim Knob control to complete the change.
6. Rotate the Trim Knob control to highlight the parameter label and
press to unhighlight.
7. Select RETURN when all changes are completed. The popup menu and
information window close.
Summary of Contents for marquette Dash 2000
Page 1: ...Dash 2000 Patient Monitor Software Version 3 Operator s Manual 2005873 001 ENG Revision C ...
Page 64: ...2 14 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Page 82: ...3 18 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Page 96: ...4 14 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Page 122: ...5 26 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Page 130: ...6 8 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Page 140: ...7 10 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Page 168: ...8 28 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Page 184: ...9 16 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Page 202: ......
Page 220: ...11 18 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...