Revision C
Dash 2000 Patient Monitor
5-5
2005873-001
If you want to move an arrhythmia call to another level for your Monitor
Default, follow this procedure. (In this example we will change V TACH
from a Crisis to a Warning level.)
NOTE: Asystole and VFIB/VTAC can only be moved when in Operating
Room Mode.
1. Rotate the Trim Knob control to move the pointer (>) up and down the
list. Stop when the pointer is in front of V TACH.
2. Press the Trim Knob control. Notice that the level for the V TACH call
is highlighted.
3. Rotate the Trim Knob control until WARNING is displayed.
4. Press the Trim Knob control to complete the change. The information
window is reorganized to include V TACH as a WARNING alarm.
The information window remains open for you to make any other
changes.
5. When you have completed all changes, move the pointer to RETURN
and press the Trim Knob control to close the information window.
Summary of Contents for marquette Dash 2000
Page 1: ...Dash 2000 Patient Monitor Software Version 3 Operator s Manual 2005873 001 ENG Revision C ...
Page 64: ...2 14 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Page 82: ...3 18 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Page 96: ...4 14 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Page 122: ...5 26 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Page 130: ...6 8 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Page 140: ...7 10 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Page 168: ...8 28 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Page 184: ...9 16 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...
Page 202: ......
Page 220: ...11 18 Dash 2000 Patient Monitor Revision C 2005873 001 For your notes ...