RFC 2544 Tabs
Ethernet and Fibre Channel Application
311
Back-to-Back
Results
Test State
--
: Indicates that the test has not run yet.
In Progress
: Indicates that the test is running.
Completed
: Indicates that the test is completed.
Aborted
: Indicates that the test has been interrupted (stopped).
Status Message
: Gives the test status. Following is the list of status
messages.
S
ending learning frames
Sending test frames
Test completed
Aborted by user
Aborted - Loss of remote connection
Link down
MAC address not resolved
Frames Count
TX-to-RX/P1-to-P2/P2-to-P1
/
Local to Remote
/
Remote to Local
: Gives
the number of frames for the indicated direction. P1-to-P2/P2-to-P1 is
only available with IQS-8510B.
TX Frame Count
: Indicates the number of transmitted frames.
RX Frame Count
: Indicates the number of received frames.
Back-to-Back Results
Frame Size
: Indicates the frame sizes used for the test. Up to seven
different frame sizes can be displayed.
Note:
The “--” indicates that the result is not available because the test has not
run yet or the measured value is not valid.
TX-to-RX/P1-to-P2/P2-to-P1
/
Local to Remote
/
Remote to Local
:
Gives the number of frames for the corresponding direction.
P1-to-P2/P2-to-P1 is only available with IQS-8510B.