Summary Tabs
Ethernet and Fibre Channel Application
127
Test Summary
On/Off
button allows enabling the test timer. An error message is
displayed and the test timer is not enabled when the provided start
time or stop time is not valid. It is not possible to enable the test timer
while the test is running. When the timer is enabled (On), it is possible
to disable it even when the test is running. This setting is disabled (Off)
by default.
When test timer is enabled, it is possible to manually stop a test case
using the main test case Stop button. However, it is not possible to start
the test case when the
Start Time
is enabled.
The test timer is automatically disabled either when the user manually
stops the test, or when the given stop time or duration has expired.
Note:
When using Visual Guardian Lite for remote control, the timer
configuration values will be based on the PC clock and not on the IQS-8500
Series and IQS-8120NGE/8130NGE. Make sure to consider the time zone
differential if it exists between the PC and the IQS-8500 Series and
IQS-8120NGE/8130NGE.
Alarm Analysis
LOC
indicates that the IQS-8510G is unable to synchronize with the
selected test clock. Available with 10Gig-E interface only.
Test Preferences
Note:
Not available with IQS-8120NGE/IQS-8130NGE and IQS-8525/IQS-8535.
Couple Start/Enable TX
allows to automatically enable the stream
transmission when the test is started. The stream transmission stops when
the test is stopped. This setting is disabled by default. Refer to
Enable TX -
On/Off button
on page 204 for more information.