Calibrate zero Screen
Figure 6-6:
10.
Press
to confirm.
The Result screen (
Figure 6-7
) displays.
Result Screen
Figure 6-7:
11.
Press
to finish the process.
12.
Allow the concentrations to stabilize and wait for two minutes after stabilization.
This screen gives a reading of the concentration of the selected gas that is present as
an impurity in the nitrogen calibration gas, as measured by the instrument. In the
example shown (
Figure 6-7
), the gas being measured is NH
3
(ammonia), and the
instrument has detected a concentration of 0.530 ppm (vol).
Gas Calibration Procedures
Installation, Operations, and Maintenance Manual
49