Removing the USB PCB
Figure 7-16:
A. USB PCB
B. Threaded spacer (part of front panel)
C. M4 flat washer
D. M4 spring washer
E. M4 nut
F. USB cable
6.
Remove and retain the three nuts (E), spring washers (D), and flat washers (C).
7.
Remove the USB PCB.
8.
Use caution when unpackaging and handling the USB PCB. Examine the
replacement USB PCB for physical damage and delamination.
CAUTION!
EQUIPMENT DAMAGE
Wear an ESD wrist strap or ground yourself to prevent damage to the USB PCB
electronics.
9.
Place the replacement USB PCB (A) on the four threaded spacers (B) that form part
of the front panel.
Troubleshooting and diagnostics
Installation, Operations, and Maintenance Manual
79