Peripheral PCA Board
Figure 7-23:
A. Peripheral PCA board
B. Gas/cell temperature monitoring harness
C. Gas/cell temperature monitoring harness
D. Gas/cell temperature monitoring harness
E. Wiring harness to cell heater relay
F. Ribbon cable to motherboard
G. Wiring harness to fan A
H. Wiring harness to fan B
I. 12 V power wiring harness to J17 on motherboard
J. M3 flat washer
K. M3 spring washer
L. M3 x 6 socket head cap screw
4.
Disconnect the ribbon cable (F) from the peripheral PCA board (A).
5.
Disconnect the two fan wiring harnesses (G and H) from the peripheral PCA.
6.
Disconnect the cell heater relay wiring harness (E)