Opening the analysis cell insulation
Figure 7-38:
A. Top insulation
B. Insulation side flap
C. Insulation end flaps
D. Insulation side flap
5.
Using a 2.5 mm Allen wrench, remove and retain the six socket head cap screws and
associated spring washers and flat washers that secure the lid on the cell; then
remove the lid.
It may be necessary to use a small flat blade screwdriver to lever the lid off.
NOTICE
The pressure inside the cell must be at atmospheric pressure in order to remove the lid.
With the lid removed, the mirrors are exposed.
6.
Clean the mirrors by applying a small quantity of methanol to a clean, dry, lens
tissue and then gently wiping the mirrors with the lens tissue.
If methanol is unavailable, IPA may be used instead; however, methanol is
preferable.
Troubleshooting and diagnostics
Installation, Operations, and Maintenance Manual
125