Removing a Laser Module
Figure 7-28:
A. Base plate
B. Pin
C. Laser module
D. Captive screw
E. Ribbon cable
F. Captive screw
8.
Inspect the replacement laser module for damage.
9.
Fit the replacement laser module in position on the base plate (A). The laser module
must mate with the two locating pins (B) on the Base Plate.
10.
Secure the laser module by tightening the two captive screws (D and F).
11.
Connect the ribbon cable (E) to the laser module (C).
12.
Remove the protective masking tape from the flanges, seal,and flamepath.
Use care not to damage the flamepath. Any damage to the flamepath will invalidate
certification.
13.
Remove lock-out tag-out labels.
14.
Apply power to the analyzer.
Troubleshooting and diagnostics
Installation, Operations, and Maintenance Manual
109