13.
Press
to cancel the manual verification.
14.
Press
to finish.
15.
Repeat steps 5-9.
16.
Press
for automatic verification.
This opens the valve (where applicable) and flows the span gas.
17.
Allow the concentrations to stabilize and wait for two minutes after stabilization.
This screen gives a reading of the concentration of the selected gas that is present,
as measured by the instrument. In the example shown in
Figure 5-12
, the gas being
measured is NH
3
(ammonia), and the instrument has detected a concentration of
2000.29 ppm.
18.
If the reading is within tolerance, no further action is required. Press
to end the
span verification process.
The display controller proceeds to the Result screen (
Figure 5-13
).
Result Screen
Figure 5-13:
19.
If the reading is outside tolerance, the instrument should be span calibrated.
Refer to
Section 6.3.2
.
20.
Press
to return to the screen shown in
Figure 5-14
.
Verifying Gas Concentrations
Installation, Operations, and Maintenance Manual
43