Replace an Ethernet switch
Figure 7-27:
A.
Ethernet connector (to analog input unit)
B.
Ethernet switch
C.
Ethernet connector (to Ethernet port)
D.
Ethernet connector (to Motherboard)
E.
Lever
F.
Wiring harness
Procedure
1.
Power down the analyzer by switching the
ON/OFF
switches on the front and rear
panel to the
OFF
position and allow the core to cool for twelve hours.
2.
Lock-out tag-out the power supply.
3.
Remove the analyzer from the rack using two people to lift the analyzer and place it
on a clean workbench.
4.
On the Ethernet switch (B), disconnect the Ethernet cable that links the Ethernet
switch to the analog input unit from Ethernet connector number 1 (A).