2
Install
2.1
Site selection
WARNING!
HEAVY ITEM
Handle the analyzer with caution during unpacking, installing, maintaining, and transporting
to prevent crushing of hands, feet, or other body parts.
The analyzer weighs 31 kg (68 lb). Emerson
™
recommends that the analyzer is only moved and
lifted by a minimum of two people. Wear suitable protective gloves and protective footwear.
Failure to properly handle the analyzer may cause injury to personnel.
Procedure
1.
Install the analyzer in a standard 482.6 mm (19 in.) rack.
NOTICE
Fitting the analyzer into the rack and securing it are your responsibilities.
2.
Install the analyzer in a suitable shelter to protect it from environmental elements
and water. The Ingress Protection rating (IP) for the analyzer is IP-44 per the IEC
60529 standard.
DANGER!
FLAMMABLE SUBSTANCES
Some parts of the analyzer may reach temperatures of 190 °C (374 °F) and may present
an ignition source. Exercise care when using oil, paint, cleaning rags, and other
flammable substances near the analyzer. A fire may result if this precaution is not
observed. Always assume that the interior of a analyzer is hot unless it has been
switched off and allowed to cool down.
Install
Installation, Operations, and Maintenance Manual
13
Summary of Contents for Rosemount CT5400
Page 15: ...Abbreviation Description WEEE Waste electrical and electronic equipment μm Micro meter ...
Page 16: ......
Page 20: ...Contents iv Rosemount CT5400 ...
Page 32: ...Plan 12 Rosemount CT5400 ...
Page 40: ...Install 20 Rosemount CT5400 ...
Page 44: ...Startup procedure 24 Rosemount CT5400 ...
Page 54: ...Operating the analyzer 34 Rosemount CT5400 ...
Page 158: ...Troubleshooting and diagnostics 138 Rosemount CT5400 ...
Page 164: ...Shutdown procedure 144 Rosemount CT5400 ...
Page 166: ...Preventative maintenance 146 Rosemount CT5400 ...